Conductivity measurement of fish and instrumentation

of swim tunnel.


 

Master of Science Physics –Industrial Instrumentation

proposal for Endre Grimsbø

 

Print friendly version in pdf-format

 

 

 

Supervisors: Associate Professor Bjørn Tore Hjertaker and Professor Ole Brix

 

Introduction

 

This master project consists of two parts

(i) instrumentation of a swim tunnel and

(ii) measurements of conductivity in living fish.

The instrumentation of the swim tunnel is implemented not only with the intention of measuring conductivity in fish, but also to make efficient use of the tunnel as a respirometer.  

The experimental results of the conductivity measurements will be used in the development of a Fish Flow Meter by Christian Michelsen Research AS (CMR). The model for conductivity measurements developed in this project is also expected to be used in upcoming projects on conductivity measurements of living fish.

The basic principle of CMR’s Fish Flow Meter is to use conductivity measurements (s) to calculate the volume of fish passing through a pipeline. Conductivity measurements is a well-developed measurement principle within multiphase flow in the oil and gas industry.  Multiphase flow is a flow consisting of different components, e.g. oil, gas and water.

 

Conductivity measurements

The use of conductivity for measurement of fish is a transfer of technology from the oil and gas industry. In order to better understand the background of the technology an introduction to the measurement methods used in the oil and gas industry is useful as a starting point.

Figure 1. Cross section of pipeline filled with two component flow consisting of oil and saline water

 

In figure 1 the two phases, crude oil and saline water, are separated and the measurement principle shown is based on measurement of electrical capacitance to determine the electrical permittivity ε of the two component flow.

In order to denote multiphase hydrocarbon flow consisting of oil, gas and water, it is common to use the following abbreviations:

α - The oil volume fraction of the total flow

β - The water volume fraction of the total flow

γ - The gas volume fraction of the total flow

 

Assuming the total volume of measurement volume is unity and that the flow only consists of oil, water and gas α+β+γ=1.

Oil and gas have low conductivity compare to water, especially saline water which has relatively high conductivity. As a result of this the previous mentioned permittivity measurement based on electrical capacitance could only performed if the oil and gas fractions of the flow prevents the saline water from short-circuiting the electrodes. Therefore, it is important to know the distribution and behavior of each component in the measurement volume. An issue when monitoring multiphase flow in a pipeline is to determine which of the different phases is continuous.

It is obvious that it is only possible to use the capacitance principle for measuring in oil or gas continuous phase. In water continuous multiphase flow the electrodes will be short-circuited due to the ions in non-distilled water. Due to the conductivity of the water component, measurements based on conductivity instead of electrical capacitance has to be used in a water continuous multiphase flow.

 

Figure 2. Flow of water with gas bubbles in a pipeline

 

A flow of water with gas bubbles in a pipeline is also called a two phase water continuous flow.

In order to calculate the mixture permittivity and conductivity of two flowing phases, expressions has been developed by e.g. Maxwell and Bruggeman. These formulae assume that the two different phases are homogeneously mixed. Since water has a relatively high conductivity, a commonly used expression is the formulae developed by Ramu and Rao.

The basis for Ramu and Rao’s equations are that σ1 << σ2 and ε1 << ε2, which is true for a two component flow contains of water and oil/gas. When this parameter is fulfilled, and the continuous phase consists of component one, then the equations are as follow:

 

             

 

In the equation σm and εm is the total the conductivity and permittivity of the mixture and b is the water volume fraction often called water cut.

 

When the continuous phase consists of component two the following equation has to be used:

 

                     

 

The tag used at σ’m and ε’m indicates that this is the total the conductivity and permittivity of the mixture where continuous phase consists of the second component and b is none continuous.

 

Connection between hydrocarbon flow and conductivity measurement in fish

The starting point was that we wanted to measure the volume of fish in a pipeline.

 

 

Figure 3.  Cross sectional area of a pipeline with a fish

 

It is now possible to look at the pipeline in figure 3 as a water continuous two component flow, and the fish is like a gas or oil bubble in a water continuous flow. By measuring the cross sectional areas it is possible to describe the total conductivity of the two components as follows:

σtot = σwater + σfish

It is now obvious that a flow of fish and water is much the same as a multiphase flow of water and oil.

Conductivity measurements of living fish has not been done previously, only measurements of dead fish. A significant part of the master project will therefore be to find a suitable model for conductivity measurement of a living organism that not necessarily wants to be measured.

                       Figure 4.  Principle of the conductivity measurement

 

The principle of the conductivity measurement that is expected to be used is named a “four-terminal conductivity meter”. The measurement principle is to send an electrical current through the water and measure the voltage drop over the voltage electrodes.   From the voltage drop it is possible to calculate the conductivity. The measurement taken does not represent the cross sectional area but the entire volume between the voltage electrodes.  Two measurements have to be taken, one with fish in the swim tunnel and one without fish. Based on the difference between the measurements it is possible to calculate the conductivity of the fish itself.

It is also necessary to calculate the volume of the fish between the electrodes. Therefore it is required to find a suitable method to calculate the volume of living fish. In previous experiments this volume measurements, related to the conductivity calculation, have only been done with dead fish. Again, the challenge probably will be to measure a fish which does not necessarily wants to be measured. Probably some kind of anesthetic has to be used in the experiment just so that the fish relaxes when the measurements are done. 

The conductivity in the water is strongly dependent on two factors, temperature and salinity. It is therefore important that the temperature is stable under the entire experiment, and an efficient temperature stabilization is needed. Temperature stabilization will be done by the flow tunnel control unit and a cooling/heating system. The cooling/heating system consists of a pump which pumps water from the flow tunnel through to a cooling/heating loop. The cooling loop has to be placed in a cooler or a heater dependent of the wanted effect.

The conductivity in seawater is normally 5 S/m and will vary with salinity. Hopefully the salinity of the seawater delivered from the supply system in the laboratory is stable. The salt water delivered by the supply system is taken from the fjord outside the laboratory, and hopefully there are no or insignificant variations in salinity.  

 

Figure 5. Conductivity in Siemens pr. meter versus salt in mg pr. liter water

 

Figure 5 is taken from a previous master thesis written by Anders Nødseth. The figure shows the relations between salinity pr mg/l in water and conductivity. Before the experiment starts it is necessary to verify that variations in salinity are insignificant and therefore can be neglected.

Figure 6. Conductivity in Siemens pr. meter versus temperature

 

Figure 6 is also taken from a previous master degree thesis written by Nils Bjarte Dyrøy. The figure shows the variations in conductivity caused by temperature variations.

In order to avoid fluctuations in the conductivity it is important to have a stable temperature. Variations in temperature will give an unwanted uncertainty in the conductivity measurements and has to be avoided.   

The control unit used in this experiment must include a control algorithm for stabilizing the temperature in the water. It is expected that an algorithm based on a simple proportional controller (P-controller) is adequate for this experiment. If the temperature stabilization is insufficient with the P-controller more advanced controllers can be used like PI or PID-controller. The control algorithm will be implemented in a Compact FieldPoint (cFP) unit used for controlling the flow tunnel and will be discussed later in this document. The programming tool used for configuration of the unit is called LabVIEW and is a graphical programming language from National Instruments.  

 

Swim tunnel as respiriometer

To get a better understanding of what a respiriomer is it is useful to take a look at gas exchange in fish in general.

Fish live in a dense and very viscous environment, which for a terrestrial animal would appear to be almost depleted of oxygen. At best, one volume of water contains a little less than 4% the amount of oxygen present in the same volume of air. In air the solubility of a gas would roughly be 1000 mL/101.3 kPa = 9.87 mL Gas.L-1.kPa-1 (at 15°C). In practice we treat the respiratory gases as ideal gases, which mean that O2 and CO2 would dissolve more or less equally in air. In water, however, the solubility of gases depends on:

·        The nature of the gas

·        The pressure of the gas in the gas phase

·        The temperature

·        The presence of other solutes

 

Thus, while the solubility of O2 in water is only about 1/30 the solubility in air, the solubility of CO2 is more or less the same in air and water depending on temperature (similar at about 15°C). For calculating gas fluxes we use a concept which gives the following three fundamental gas equations:



is the volume of gas taken up (O2) or produced/released (CO2). is the flow of medium over the gills.  is the capacitance coefficient (solubility) of the gas in the medium. P is the partial pressure of gas (i: before gills; e: after gills). From these expressions, we may calculate:


RQ is the respiratory quotient (R for exchange ratio is used when examining isolated parts of the fish). Rearranging the formulas and ignoring Pi we get:


If the fish extracts all the oxygen in the water (Pi @ 21 kPa, Pe = 0 kPa), the CO2 partial pressure can, at a maximum, be about 0.7 kPa in the exhalent water. This means the maximum partial pressure in the arterial (post branchial) blood will not exceed this value, which is almost a factor 10 lower than for air breathing animals.

This low oxygen availability has undoubtedly contributed to the evolutionary development of:

·        Large gill surface areas for extremely efficient gas exchange.

·        Developments of air breathing organs and the necessary circulatory arrangements.

·        Acid-base regulation depending more on ion than ventilatory exchanges.

To measure the gas exchange in fish it is common to use a respiriometer. There are several different types of respiriometers based on different principles or systems.

 

                          

 

Figure 7 Closed system respiriometer.

 

Let’s have a look at the type called closed system, shown in figure 7. The construction may vary depending on the material used. There may either a continuous registration of oxygen partial pressure by an internal sensor or more primitively water may be sampled (and replaced) for O2 external analysis. It is important that the sensor is kept at the same temperature as the fish. If the fish is sluggish or we measure the consumption of fish eggs O2 gradients around the fish or eggs may build up, and we need to stir the water either by a magnetic stirrer or a small pump (the later can be combined with an external sensor). We also need to optimize the size of the chamber to prevent long experimental runs with building up of organic waist products from the fish, which may influence the metabolic rate (i.e. carbon dioxide). For measuring the metabolism gas tight glass syringes have often been used. The closed principle is also used in activity studies in the so called tunnel respirometer.

Due to the measuring principle in the closed system we will never be able to measure the oxygen consumption at a constant oxygen partial pressure. In order to obtain a proper precision of measurement we need a decrease in oxygen partial pressure of minimum 0.7 kPa. However, this technique is very useful to obtain the critical oxygen tension, Pc, at which the fish is no longer able to regulate its oxygen consumption in relation to the decrease in ambient oxygen partial pressure (see figure next page).

 

 

Figure 8. The different types of metabolism in relation to environmental oxygen partial pressure.

 

The method is very simple and the precision of measurement is very high. At the same time we can reduce the response time to a minimum. In addition to the problems regarding accumulation of organic waist products we also have to take into account the metabolic state of the fish, since the measurements starts immediately after transferring the fish to the respirometer. We will thus never obtain the so called standard metabolism but rather a more active metabolism which we call the routine metabolism.

 

 

Figure 9. The open respirometer

 

The open – or flow through system. This system has been preferred by many because it is easy to control the water quality. This feature also allows for acclimation of the fish prior to measurements allowing it to calm down. As for the closed system the respirometer consist of a gas tight Perspex housing. There is a whole in both ends allowing a constant flow of water through the respirometer. The flow is either controlled by a leveling system or by a pump. The only draw back of the system is a ‘time lack’ of response caused by a combination of flow, chamber size, metabolism and oxygen gradients in the system. As a result we don’t get a 100% registration of all changes in metabolism at any time. On the other hand we are able to do long term runs without accumulation products and with full control of physical parameters in the water.

For both systems it’s important that we use a homogeneous material that has been acclimated to the proper condition without feeding 24 hr prior to experiment. Further more any disturbances of the fish should be avoided during experiment.

 

Figure 10 Stop-flow respiriometer.

 

Finally the principle used in this experiment called the stop-flow system, shown in figure 10. This system is a combination of the open flow system and the closed system. At the figure there is a closed circulatory system with a propeller which stirrers the water. It is a electric motor that rotates the propeller. The electric motor get it’s power supply from a adjustable frequency AC drive that is controlled by the swim tunnel’s main controller unit.

The closed circulatory system, or inner camber, consists of flow tunnel and a camber where the fish swim. (position 2 and 3 at the figure)

An optode for oxygen measurement and temperature sensor is mounted in flow tunnel and connected to an instrument who communicate whit the main controller unit. Position 1 at the figure is the outer chamber or reservoir.  An open system is connected to the reservoir, where water quality and physical parameters in the water are controlled. Water oxygen partial pressure is controlled by a feedback control using magnetic valves connected to air compressor/oxygen tank and a nitrogen tank. When the flush pump is running, water from reservoir is pumped in to inner camber. Water from inner camber is now flushed out the camber and replaced with oxygen rich water. Figure 10 also shows the temperature regulation pump that pumps water through the cooler/heater depended on the temperature wanted. All the actuators, i.e. pumps, electric motor and magnetic valves are controlled by the main controlling unit.

 

           Figure 11. Oxygen partial pressure when running the respiriomerer.

 

The oxygen level increase when the flush pump starts and flushing the system with water from the reservoir. (Ascending curve) The flush pump stops i.e. switching to a closed system and wait to brake down oxygen gradients. Then a descending straight line showing the decrease in oxygen partial pressure as a result of the oxygen consumption of the fish measured. Let’s now take a look at the swim tunnel before discussing the main control system.

 

Figure 12 the swim tunnel seen from above, with all the equipment

including the electrodes for conductivity measurement.

 

The swim tunnels main control device named compact FieldPoint (cFP) and is manufactured by National Instrument (NI). The cFP is described by the producer as a programmable automation controller (PAC). A PAC is a combination of a personal computer (PC) and a programmable logic controller (PLC). The common used Norwegian name of a PLC is “programmerbar logisk styring” (PLS).

The cFP look like and acts pretty much like a module based PLC with controller and I/O modules mounted on a back plane.

 

Figure 13. A cFP with controller and I/O modules

 mounted on a backplane. (from NI’s web site)

 

The I/O modules to control the swim tunnel are a relay module, an analog output module and a possibly a digital output module. The relay module controls the flush pump, cooling pump, oxygen and nitrogen valve connected to the swim tunnel. Connections of devices to the relay module implement a “fail to safe” condition, i.e. the flush pump is running and the oxygen valve is open. The analog output module is intended to control the adjustable frequency AC drive and thereby the flow in swim tunnels inner chamber. Measurements of oxygen and temperature are done by an optode and a sensor, previous mentioned, these are connected to instrument named Microx TX3. The Microx TX3 instrument is connected and communicates with cFP through the RS232 communication protocol.

The instrument for measuring conductivity in fish, the 3522-50 LCR HiTESTER, will be connected to the electrodes placed in the swim tunnel. The conductivity instrument also communicates with the cFP through the RS232 communication protocol.

The controller and its interface is the significant difference between a regular PLC and a cFP. In fact the controller unit on a cFP is a small PC with a built-in web server and support for file transfer protocol (ftp). The cFP communicates over a regular Ethernet and connects to the network at UiB with its own IP-address. The network properties make the cFP, a suitable device in a so called distributed control system (DCS), i.e. a control system with several control devices communicating over a network.

A special real-time operating system is running the controller and not a common PC operating system like Windows or Linux. Because of the characteristic of the cFPs operating system the device is also called a real-time controller (RTC).   

The programming of cFP has to be done with NI’s development tool named LabVIEW and a dedicated real-time (RT) software module.  A control program with necessary functionality to measure conductivity and run the swim tunnel as a respirometer has to be developed. The developed program must be compiled as an executable file, i.e. a so called exe-file, so it can be used by others without any knowledge of LabVIEW.

It is important to note that the LabVIEW runtime engine needs to be installed on users PC before it is possible to run the exe-file. To make a simplified picture of the situation the runtime engine is a software layer between the operating system and the exe-file. The concept with LabVIEW runtime engine is pretty much the same as JAVAs virtual machine or Microsoft .NET-framework.

 

 

 

Figure 14. Arrangement of the swim tunnels control system

 

When the exe-file is executed it gives the system a human machine interface (HMI) at the laptop/PC, that gives the operator access to the system.

It is also needed to compile and store a version of the exe-file in the RTC’s memory as an embedded application. When the PAC is connected to a power supply and switched on, the embedded application has to be executed under the RTC’s boot up sequence.

The PAC will, because of the embedded application, have the ability to be a stand-alone device if the connection to other DCS devices fails. As a stand-alone device the PAC has only some LED’s and opportunity for some switches as the only HMI and the user will have little or almost no access to the system.

As an option it is possible to use one switch and a diode to make a calibration procedure for the Microx TX3 that is connected to the cFP.

 

Figure 15. The cFP unit connected in the network,

 including other network devices needed for the experiment

 

The web cameras are needed in order to observe the fish during the experiment. During the conductivity experiments it is important to know exactly where the fish is located compared to the placement of the electrodes. It is also an option to make an automatic recognition to determine the behavior of the fish.

The measurements taken will be stored on a server in the network, named “data” in the figure.

If possible it is interesting if the experience with the swim tunnel and it’s equipment can result in suggestions for improvement of future swim tunnels or experiments. It is also of interest if this master project may lead to new ideas and technology. 

 